Analytical balance
ALJ 200-5DA

Product family: ALJ

Analytical balance with high weighing capacity, graphic display and convenient recipe function

Technical Data

Weighing capacity [Max] 82 g; 220 g
Readability [d] 0,00001 g; 0,0001 g
Units g, mg, gn, dwt, tl (Tw), tl (HK), ozt, tl (Singap, Malays), ct, mo, lb, oz
Linearity ± 0,0001 g; 0,0002 g
Reproducibility 0,00004 g; 0,0001 g
Stabilization time 10 s; 6 s
USP Minimum weight of sample (k = 2, U = 0.1%) 80 mg
Interfaces RS-232 standard
Adjustment options Internal calibration (automatic)

more data

Your price
€ 2.250,00 *
*plus shipping costs (not applicable for downloads), plus VAT
Available from stock
Compare Technical data sheet Product brochure Download centre

Do you have questions?

We will be happy to consult you on our KERN scales and test weights

Tel.: +49 7433 9933-560

scales@kern-sohn.com

Standard

Automatic internal adjustment
RS 232 data interface
GLP/ISO Protocol
Piece counting
Recipe level B
Percentage determination
Weighing units
Weighing with tolerance range, visual, acoustic
Function_Mean value_Animal
Universal mains adapter_C
Weighing principle: Force compensation
Delivery_Express_1_day

Option

Certificate_DAkkS_3_days

Description

  • Rapid and efficient operation thanks to the graphics display. Simple, clear user interface on the display in the following languages: DE, EN, FR, IT, ES, PT
  • Automatic internal adjustment, guarantees high degree of accuracy and makes the balance independent of its location of use
  • Short stabilisation time: steady weight values within approx. 10 s; 6 s under laboratory conditions
  • Weighing with tolerance range (checkweighing): Input of an upper/lower limit value. A visual signal assists with portioning, dispensing or grading
  • Dosage aid: High stability mode and other filter settings can be selected
  • Internal memory for complete recipes with name and target value of the recipe ingredients. User guidance through display
  • Ergonomically optimised keypad for left and righthanded users
  • Large glass draught shield with 3 sliding doors for easy access to the items being weighed, Weighing space W×D×H 160×170×225 mm
  • Compact size, practical for small spaces
  • Protective working cover included with delivery

Note

  • The USP minimum sample weight is a measure for the smallest permitted weight in accordance with USP (United States Pharmacopeia).
  • ALJ-A03: Ionizer to neutralise electrostatic charge for fixed installation in the analytical balance. Particularly convenient handling as you no longer need a separate device. Simply enable the ionizer fan at the push of a button. Suitable for all models, see accessories

Technical Data

Weighing capacity [Max] 82 g; 220 g
Readability [d] 0,00001 g; 0,0001 g
Units g, mg, gn, dwt, tl (Tw), tl (HK), ozt, tl (Singap, Malays), ct, mo, lb, oz
Linearity ± 0,0001 g; 0,0002 g
Reproducibility 0,00004 g; 0,0001 g
Stabilization time 10 s; 6 s
USP Minimum weight of sample (k = 2, U = 0.1%) 80 mg
Interfaces RS-232 standard
Adjustment options Internal calibration (automatic)
Ambient temperature 5 °C – 35 °C
Dimensions weighing surface (Ø) 80 mm
Material draught shield glass
Material housing plastic
Dimension housing (W×D×H) 210×340×330 mm
Material weighing plate stainless steel
Net weight approx. 7 kg
Construction type of the scale Multi-range balance

Accessories

Software

Services

Spare part wanted?

Here you will find the right spare part for this product: